LINQ II

Model LNQ22 | Medtronic, Inc. | ICM

Models with the same MRI conditions

LNQ22

The MRI conditions shown are based on the manufacturer's official guidelines.
Do not apply them to models not listed above; search for that model separately. Search models →

Important alert

  • Required: Ensure that the patient has no other devices, leads, or implanted items that are known to pose a hazard in an MR environment.

1Pre-scan checks

    • That the patient has a LINQ II device.
    • That the patient has no other devices, leads, or implanted items that are known to pose a hazard in an MR environment.
    • Ensure that the pre-scan cardiology responsibilities have been performed.
    • You can use an x-ray image of the LINQ II device, such as shown in Figure 1, to identify that the patient has a LINQ II device. LINQ II devices are typically implanted in the left chest area. [p. 6]
    • Required programming settings: None prior to the MRI exam.
    • If you have questions about whether the patient should receive an MRI scan, contact the patient’s cardiologist.

    2MRI scan conditions

    ≤3 T

    Field strengthStatic magnetic field (B0) [T]: ≤3 T
    Scanner typeType of nuclei: Hydrogen proton
    Static magnetic field (B0) orientation: Horizontal, cylindrical bore / Perpendicular to patient, AP (anterior – posterior) / Perpendicular to patient, LR (left – right)
    Spatial gradientMaximum spatial field gradient (SFG) [T/m and gauss/cm]: ≤25 T/m (2500 gauss/cm)
    Gradient slew rateMaximum gradient slew rate per axis (T/m/s): ≤200 T/m/s
    Operating modeRF operating mode or constraints: First level controlled operating mode or normal operating mode
    Whole-body SARWhole-body specific absorption rate (SAR) [W/kg]: ≤4 W/kg
    Head SARHead SAR [W/kg]: ≤3.2 W/kg
    B1+rmsMaximum B1+RMS [µT]: No limit (see entry for Whole-body specific absorption rate)
    Scan region / position conditions• Anatomy at isocenter: No patient-landmark restrictions.• Patient position in scanner: Any• Scan duration and wait time: No limits on scan duration or post-scan wait time for the labeled RF conditions. No limits on post-insertion wait time prior to an MRI scan for the labeled RF conditions.

    3RF COIL / RF excitation

    RF coil• RF transmit coil type: Integrated whole-body transmit RF coil / Detachable head transmit/receive RF coil / Detachable extremity transmit/receive RF coil• RF receive coil: Any receive-only RF coil may be used.

    4Other information

    • It is acceptable to scan patients who have multiple MR Conditional devices, as long as the MR labeling conditions for all implants are satisfied.
    • It is acceptable to scan patients who have LINQ II devices that have been abandoned if the MRI conditions for use are satisfied.
    • Do not bring the tablet that the Reveal LINQ Mobile Manager clinician app is installed on, the Patient Assistant, or the patient monitor into Zone 4 (MRI magnet room), as defined by the American College of Radiology. They are MR Unsafe.
    • Image artifact and distortion resulting from the presence of the device within the field of view must be considered when selecting the field of view and imaging parameters. These factors must also be considered when interpreting the MRI images.
    • Due to the static magnetic field and gradient magnetic fields produced by MRI equipment, the magnetic material of an implanted device may exert force, vibration, and torque effects that the patient may or may not feel. Such effects will not harm the patient or damage the device.

    These conditions are based on the source document shown below (revision and date). Before scanning, verify the patient's complete system (device, leads, and any other implants) against the latest manufacturer labeling.

    5Source documents

    Source Title: LINQ II™ Insertable Cardiac Monitor — MRI procedural information — MRI Technical Manual

    Document No.: M028264C001 B

    Revision / Date: B / 2022-10-11

    Official URL: www.medtronic.com/manualsSearch for document number M028264C001 on this site.